IEC 62304 Software Lifecycle for Medical Devices: A Practical Guide

IEC 62304 is the lifecycle standard behind most software submissions. This guide explains the safety classes, the processes you must run and document, how SOUP is handled, and what to expect from Edition 2.

Reviewed by Aakash Jindal, Founder, iMDR MedTech. Regulatory status is checked and dated on each guide.

Classification

Software Safety Classes (Edition 1.1)

ClassPossible harm from a software failureEffect on the work
Class ANo injury or damage to healthLightest documentation: development plan, requirements, system testing, release and basic maintenance records
Class BNon-serious injuryAdds architecture and detailed design records, and verification of software units where appropriate
Class CDeath or serious injuryFull set: detailed design, unit verification, and the most rigorous integration and testing

The class is assigned from the risk analysis, and the standard allows a segregated item to carry a lower class when the segregation is justified. Confirm the rules in the edition you apply.

Lifecycle

The Processes You Run and Document

Software development planning

Define processes, deliverables, tools, traceability, configuration management and how SOUP and problems will be handled.

Requirements analysis

Capture functional, performance, interface and risk control requirements, and verify them.

Architectural and detailed design

Break the system into items, define interfaces and identify SOUP. Class B and C add more detailed design records.

Implementation and unit verification

Write the code and verify units to the level the class requires.

Integration and integration testing

Combine the items and test the interfaces and the integrated whole.

System testing and release

Test against the requirements, resolve anomalies, and release only with a controlled baseline and documented known issues.

Maintenance and problem resolution

Run a change process that assesses each change for risk and regression, and keep a traceable problem log.

Making It Work With Agile and Fast Releases

IEC 62304 describes outcomes, not a waterfall sequence, so iterative teams can comply. What matters is that the plan states how your process produces the records: requirements that trace to tests, design reviews that happen, and a release baseline that is controlled. Automate traceability where you can, and keep the plan short enough that the team follows it.

Linking to Risk Management

The software safety class comes from the risk analysis, and risk controls implemented in software become requirements you must verify. Keep the software risk file and the device risk file consistent. See risk management and usability.

Edition 2: What We Know and What We Do Not

When last checked on 28 November 2026, Edition 2 had not been published. Sources disagreed on whether it was at final-draft stage, and forecasts for publication ran from late 2026 to mid-2027. Draft reports describe a wider scope that covers health software generally, a change to how classes are expressed, and added material on AI-enabled software, but the text can still change. Regulators typically take time to recognise or harmonise a new edition, so the current edition remains the working basis. Keep a clean mapping of your records to the standard so that a later gap review is quick.

FAQ

Frequently Asked Questions

What is IEC 62304?

An international standard for the software lifecycle of medical devices. It defines the processes for planning, requirements, architecture, design, implementation, verification, release, maintenance, risk management, configuration management and problem resolution. FDA and EU regulators expect software to follow a controlled lifecycle, and this is the standard most teams use to show it.

What are the software safety classes?

Edition 1.1 has three classes based on the harm that software failure could contribute to. Class A: no injury or damage to health is possible. Class B: non-serious injury is possible. Class C: death or serious injury is possible. The class sets how much documentation and verification you need.

Is IEC 62304 Edition 2 published?

Not when this page was last checked. Edition 2 was still in development, and sources disagreed on its stage and publication date. Drafts reported a wider scope covering health software generally. Build to the current edition and monitor the standard.

What is SOUP?

Software of unknown provenance: software you did not develop for your device, such as open-source libraries, operating systems and third-party components. IEC 62304 requires you to specify, evaluate and manage the risks of each item, including known anomalies and updates.

Does IEC 62304 cover cybersecurity?

Not in depth. Cybersecurity is addressed alongside it, for example through FDA premarket cybersecurity expectations and security standards such as IEC 81001-5-1. Plan both together.