EUDAMED: What Is Mandatory Now and What Manufacturers Must Do

Four EUDAMED modules became mandatory on 28 May 2026, and the deadline for registering devices already on the market falls on 28 November 2026. This guide sets out the dates, who must act, and what to prepare.

Reviewed by Aakash Jindal, Founder, iMDR MedTech. Regulatory status is checked and dated on each guide.

Key Dates

The EUDAMED Timeline

DateWhat appliesWho acts
27 November 2025Commission Decision (EU) 2025/2371 published in the Official Journal, declaring four modules functional and starting a six-month period under Regulation (EU) 2024/1860No action needed
28 May 2026Mandatory use of the actor registration, UDI and device registration, notified bodies and certificates, and market surveillance modules. New devices must be registered before they are first placed on the marketManufacturers, authorised representatives, importers, notified bodies, authorities
28 November 2026Deadline to register devices that were placed on the EU market before 28 May 2026 and are still being suppliedManufacturers (through their registrations)
28 May 2027Information on certificates issued before 28 May 2026 must be enteredNotified bodies, with manufacturer data
Not yet setVigilance and post-market surveillance, and clinical investigations and performance studies. Mandatory use follows a Commission notice that each module is functionalWatch the Commission’s EUDAMED page

Status checked 28 November 2026 against the European Commission, the Irish regulator HPRA and notified body guidance. Confirm current dates on the Commission’s EUDAMED page.

Action Plan

What Manufacturers Should Do

Confirm every economic operator is registered

Manufacturer, authorised representative and importers each need an actor registration and an SRN. Without a valid SRN, you cannot register devices or have certificates linked to them.

List your devices and group them correctly

Decide the Basic UDI-DI groupings first, then the UDI-DIs beneath them. Mistakes in grouping are hard to unwind once certificates and declarations point at them.

Collect the device data in one controlled sheet

Risk class, intended purpose, applicable regulation, notified body and certificate details, and labelling attributes. Take them from the technical documentation so EUDAMED, the label and the file agree.

Register new devices before first placement

There is no grace period for new devices. Build registration into the release checklist, next to the declaration of conformity and the label approval.

Triage the legacy portfolio

For devices already on the market, map the existing UDI-DIs and prepare a bulk upload. A device family may not need its own entry where an equivalent MDR or IVDR device is already registered, but record the reasoning.

Coordinate with your notified body

Certificates can only be linked to devices that are already registered. Agree who enters what, and check that your certificate details match what the notified body uploads.

Common Mistakes to Avoid

Treating registration as paperwork for later. A device that is not in EUDAMED should not be placed on the market, so the registration is a release gate and belongs on the release checklist.

Registering from memory. Data entered from a spreadsheet that nobody has reconciled with the technical documentation creates three versions of the truth. Reconcile first.

Assuming the authorised representative handles it all. Your authorised representative may register the actor entry for a non-EU manufacturer, but the device data is yours. Agree the split in writing.

Forgetting the second market. EUDAMED covers the EU. Other markets keep their own registers, and the US uses GUDID. Consistent device identifiers across both save rework. See the FDA vs EU MDR comparison.

Set Expectations

What EUDAMED Registration Does Not Do

  • Approve your device — Registration is not conformity assessment. CE marking still depends on the technical documentation, the quality system and, where required, a notified body certificate.
  • Replace vigilance reporting rules — Until the vigilance module is mandatory, follow the reporting routes your competent authorities require.
  • Fix an inconsistent file — Data in EUDAMED must match the label, declaration of conformity and technical documentation. Conflicts surface during audits and market surveillance.
FAQ

Frequently Asked Questions

Is EUDAMED mandatory?

In part. Four modules have been mandatory since 28 May 2026: actor registration, UDI and device registration, notified bodies and certificates, and market surveillance. The remaining modules, vigilance and post-market surveillance, and clinical investigations and performance studies, had not been declared functional when this page was last checked.

Do I need a Single Registration Number (SRN)?

Yes, if you are a manufacturer, authorised representative or importer placing devices on the EU market. Economic operators register in the actor module and the competent authority validates the entry and issues the SRN. Manufacturers outside the EU work through their EU authorised representative.

Do devices already on the market have to be registered?

Yes, with a longer deadline. Devices placed on the EU market before 28 May 2026 and still sold afterwards must be registered by 28 November 2026. New devices must be registered before they are first placed on the market, with no transition period.

What is a Basic UDI-DI and where do I find it?

The Basic UDI-DI is the main key that identifies a device model or family in EUDAMED and links its data, certificates and declaration of conformity. It is not printed on the label. The UDI-DI on the label identifies a specific packaging level and version.

Does EUDAMED apply to in vitro diagnostics too?

Yes. EUDAMED supports both the MDR and the IVDR, and the four mandatory modules apply to both.

Need Help Registering Devices in EUDAMED?

Describe your portfolio and deadline. Matching experts in EU MDR and IVDR registration are put forward after team review.