What a PSUR Must Contain
Article 86(1) requires a summary of the results and conclusions of the analysis of post-market surveillance data gathered under the plan, together with a description of any preventive and corrective actions taken and the reasons for them. Over the life of the device it must also state:
- the conclusions of the benefit-risk determination;
- the main findings of the post-market clinical follow-up; and
- the sales volume of the device, an estimate of the size and other characteristics of the population using it and, where practicable, how often it is used.
MDCG 2022-21 adds a recommended template, a PSUR reference number that stays the same through the life of the report, examples of how to present data, and a web form with the data to be stored in EUDAMED for class III and implantable devices. Read it before you design your own template.
Common Mistakes to Avoid
Restating last year’s report. Each update should reflect what happened since the previous one. Reports that repeat the same risk wording with new dates are an easy audit finding.
A plan with no thresholds. If the plan does not say what level of complaints, trend or literature signal triggers action, the PSUR cannot show that you acted on it.
Sales data that does not match the device grouping. The sales figures and user estimates must cover the same devices as the report. Fix the grouping first.
Forgetting the clock. MDCG 2022-21 sets how the annual and two-year cycles are counted, including for legacy devices. Put the next due date in your quality system and review it at each management review.
Treating the PSUR as separate from the CER. The clinical evaluation, PMCF report and PSUR should tell one consistent story about the device.