SaMD Guide

FDA Pathways for Software: 510(k), De Novo or PMA

The pathway you choose sets your evidence, your timeline and your cost. This guide covers how US reviewers think about software, what to settle before you submit, and what you owe FDA after clearance.

Reviewed by Aakash Jindal, Founder, iMDR MedTech. Regulatory status is checked and dated on each guide.

The Choice

Three Pathways Compared

PathwayWhen it fitsPoints to know
510(k)A legally marketed predicate with the same intended use, and no new questions of safety and effectivenessSubstantial equivalence is the test, so the predicate you choose shapes the whole submission. eSTAR is required.
De NovoA novel device of low to moderate risk with no suitable predicateIf granted, the device can serve as a predicate for later 510(k)s. eSTAR has been required since 1 October 2025.
PMAClass III, high-risk devicesNeeds valid scientific evidence of safety and effectiveness, typically including clinical data, and is the longest route.
Not regulated or enforcement discretionFunctions outside the device definition, or low-risk functions FDA has chosen not to enforce againstEnforcement discretion is a policy that can change. Record the reasoning and re-check it.

Orientation only. FDA assigns classification and product codes, and a Pre-Submission or 513(g) request can confirm them.

Predicate Strategy

For a 510(k), the predicate decision comes first. Search FDA’s public 510(k) database and, for machine-learning products, FDA’s published list of AI-enabled devices to see what has been cleared and how it was described. A predicate must share your intended use. Technological differences are acceptable only if they do not raise new questions of safety and effectiveness. A single phrase in your indications for use, such as a new clinical claim or user group, can remove a predicate and move you to a De Novo request, so settle the wording before you settle the plan.

Readiness

What to Settle Before You Submit

Ask FDA the specific questions

Use a Pre-Submission for questions that change your plan, such as the predicate, the testing approach or the AI change plan. Send focused questions, not a general request for approval.

Choose the software documentation level

FDA’s premarket guidance for device software functions sets documentation expectations by risk. Decide the level early, because it determines the design, risk and testing records you must produce.

Build the cybersecurity file if you are a cyber device

Section 524B of the FD&C Act requires a plan to monitor and address vulnerabilities, patching processes and a software bill of materials, among other items.

Plan verification, validation and performance testing

Match the test evidence to the claims. For algorithms, document datasets, independence of test data and performance across relevant groups.

Prepare in eSTAR

Fill the template as you work rather than at the end, and keep attachments consistent with what the template states.

Plan for questions from the reviewer

Additional-information requests are common. The sponsor typically has up to 180 days to respond, and the review clock stops meanwhile.

After Clearance

What You Owe FDA Once You Are on the Market

DutyNote
Establishment registration and device listingRequired for manufacturers; foreign manufacturers also name a US agent.
Unique Device Identification and labellingLabels and database entries must match the cleared device.
Quality management system21 CFR Part 820, the Quality Management System Regulation, has been in effect since 2 February 2026 and incorporates ISO 13485 by reference.
Complaints and adverse event reportingComplaint handling and medical device reporting run from the day of first sale.
Software changesAssess each change against FDA’s software change guidance, or work within an authorised PCCP. See the PCCP guide.

Check the current regulation and FDA guidance before relying on this summary.

Set Expectations

What Clearance Does Not Give You

  • Reimbursement or coverage — FDA authorisation does not decide whether anyone pays for the software.
  • Privacy compliance — HIPAA, state privacy laws and FTC health-claims rules sit outside FDA clearance.
  • Other markets — EU and other regions need their own routes; see the FDA vs EU MDR comparison.
  • A promised date — No expert can guarantee a clearance or a timeline, because both depend on the regulator.
FAQ

Frequently Asked Questions

Should software use a 510(k) or a De Novo request?

A 510(k) fits when a legally marketed predicate has the same intended use and your differences do not raise new questions of safety and effectiveness. A De Novo request fits a novel low-to-moderate risk device with no suitable predicate. The wording of the indications for use often decides which applies.

Is a Pre-Submission required?

No, it is optional. Many software teams use one, especially for novel intended uses, AI features or unusual testing, because it puts specific questions to FDA before you invest in the submission. FDA feedback is non-binding, but it is the best signal of what reviewers will expect.

What is eSTAR and is it mandatory?

eSTAR is FDA’s electronic submission template. It has been required for 510(k) submissions since 1 October 2023 and for De Novo requests since 1 October 2025, unless exempted. FDA has said it does not intend to run a refuse-to-accept review on eSTAR submissions, though it screens them technically.

How long does FDA take to decide?

FDA review goals are counted in FDA days, and the clock stops while you answer additional-information requests, so elapsed time is longer than the goal. Plan on total time, and ask an expert with recent experience of your product type.