EU MDR Device Classification: The 22 Rules Explained

Class decides your conformity assessment route, your clinical evidence and your cost. This guide explains how the 22 rules are organised, how to apply them in order, and where classifications most often go wrong.

Reviewed by Aakash Jindal, Founder, iMDR MedTech. Regulatory status is checked and dated on each guide.

The Classes

Four Classes and What They Mean

ClassRisk levelConformity assessment
Class ILowManufacturer declares conformity. Sub-groups Is, Im and Ir need a notified body for sterility, measuring function or reprocessing aspects
Class IIaMediumNotified body involvement, with sampling of technical documentation
Class IIbMedium to highNotified body involvement, with assessment of technical documentation for each device group; some implantable and active drug-delivering devices follow extra review
Class IIIHighNotified body assessment of technical documentation for each device, with extra review of the clinical evaluation for certain devices

A summary for orientation. Annex IX to XI set the actual routes, and the notified body confirms what applies.

Annex VIII

The 22 Rules in Four Groups

Rules 1 to 4: Non-invasive devices

Default class I, with higher classes for devices that channel or store substances for the body, that change the composition of blood or body liquids, or that touch injured skin or mucous membrane.

Rules 5 to 8: Invasive devices

Classes depend on the body site (orifice or surgical), the duration (transient, short-term, long-term) and whether the device is implantable or touches the heart or central nervous system.

Rules 9 to 13: Active devices

Energy delivery, diagnostic and monitoring devices, software (Rule 11), devices that administer or remove medicines, and a class I catch-all for other active devices.

Rules 14 to 22: Special rules

Devices with a medicinal substance, contraception and sexually transmitted infection prevention, disinfectants, X-ray image recorders, animal or human tissue, nanomaterials, inhaled medicines, absorbed substances and closed-loop systems.

Method

How to Classify a Device

Fix the intended purpose first

Classification follows what you claim the device is for, as stated in labelling, instructions and promotional material. Write it in one precise sentence before you touch the rules.

Apply the definitions

Check the Annex VIII definitions of duration, invasiveness, active device, reusable surgical instrument and critical anatomical areas. Many wrong classes come from skipping this step.

Run every rule, not the first that fits

List every rule that could apply to the device and to each accessory. The highest class wins.

Classify accessories and parts separately

An accessory is classified in its own right, and not by the class of the device it is used with.

Write the rationale

Record each rule considered, why it applies or does not, and the sources used. This goes into the technical documentation and is what the notified body reads.

Check against MDCG 2021-24 and agree it early

Compare with the worked examples, then confirm the class with your notified body before you invest in the evidence.

Where Classifications Go Wrong

Reading the rules by device type. The rules ask about purpose, contact, duration and energy, not about product names. Two similar-looking products can fall in different classes.

Missing the software rule. Software that informs decisions about diagnosis or therapy is classified under Rule 11 and often lands higher than expected. See the Rule 11 guide.

Ignoring the special rules. A device with a medicinal substance, nanomaterial or animal-origin tissue is caught by a special rule whatever else it does.

Copying the US class. FDA and the EU use different logic. Qualify and classify for each market on its own. See the FDA vs EU MDR comparison.

Treating the class as permanent. A new claim, a new site of use or a software change can move the class. Re-check on every significant change. The Commission’s 2025 revision proposal for the MDR was still pending at the last check, so confirm the rules in force.

FAQ

Frequently Asked Questions

How many classification rules does the EU MDR have?

Annex VIII of Regulation (EU) 2017/745 has 22 rules, up from 18 under the old directive. They sit in four groups: non-invasive devices, invasive devices, active devices and special rules. If more than one rule applies, the highest class results.

What are the EU device classes?

Class I, class IIa, class IIb and class III, rising with risk. Class I has three sub-groups that need a notified body for part of the work: Is (sterile), Im (measuring function) and Ir (reusable surgical instruments).

Does a device keep its class when it is sold in the US?

Not necessarily. FDA uses classes I, II and III under a different system. Classify separately for each market and do not carry a class across.

Who decides if there is a disagreement over the class?

The manufacturer proposes the class and the notified body assesses it. If they disagree, the matter can be referred to the competent authority of the Member State where the manufacturer is registered.

Where can I find worked examples?

The Medical Device Coordination Group guidance MDCG 2021-24 on classification explains the rules with examples. Use it alongside Annex VIII.

Unsure of Your Device Class?

Share the intended purpose and stage. Matching EU MDR classification experts are put forward after team review.