FDA 510(k) for Non-Software Devices: Predicates, Testing, Fees and Timeline

A 510(k) is a comparison. You show FDA that your device is as safe and effective as one already on the market. This guide covers how to choose the predicate, which testing to plan, what the review clock looks like and what the fee is this fiscal year.

Reviewed by Aakash Jindal, Founder, iMDR MedTech. Regulatory status is checked and dated on each guide.

The Review Clock

What Happens After You Submit

FDA dayWhat happensWhat you do
Day 1FDA receives the submission. The 510(k) must be an eSTAR submission unless exempted, and the fee must be paidHave the user fee payment and the eSTAR ready before you send
By day 7FDA sends an acknowledgement, or a hold letter if there are unresolved fee or format issuesFix any fee or format problem quickly
By day 15Acceptance review, which for eSTAR is a technical screening. FDA tells you whether the submission is accepted for substantive review or placed on holdRespond to any acceptance hold with a complete answer
By day 60Substantive interaction. FDA either proceeds with interactive review or sends an additional information request that places the submission on holdAnswer additional information requests fully, within the response window (up to 180 days)
By day 90 (FDA days)FDA’s goal for a decision on most 510(k)s. FDA days exclude time on holdPlan for elapsed time, not only FDA days

Status checked 28 November 2026 against FDA’s 510(k) submission process page and MDUFA V performance goals. MDUFA V runs to 30 September 2027. Timelines are goals, not guarantees.

Budget

FDA User Fees for FY 2027

FeeStandardSmall business
510(k) premarket notification$28,653$7,163
Annual establishment registration$13,785$13,785 (the small business discount does not apply)
510(k) reviewed by an FDA-accredited third partyNo FDA 510(k) fee, though the reviewer may charge its own feeSame

Applies to the fiscal year from 1 October 2026 to 30 September 2027, as published by FDA in the Federal Register in July 2026. Small business status must be requested each fiscal year before you file. A filing that crosses 1 October moves to the next year’s fee.

Build the Submission

From Classification to Clearance

Confirm the route

Find the classification regulation and product code for your device, check whether it is exempt, and confirm that a 510(k) rather than a De Novo request fits. A 513(g) request or a Pre-Submission can confirm it. See the FDA pathway guide for the wider choice.

Choose the predicate with care

A predicate must be legally marketed and share your intended use. Search FDA’s 510(k) database for the product code and read the summaries of the closest candidates. Check the indications for use word by word.

Build the comparison table

Compare intended use, design, materials, energy source, performance and labelling with the predicate. Where you differ, explain why the difference does not raise new questions of safety and effectiveness and show the data.

Plan the testing against the device

Use the table below, FDA guidance and any special controls for the product code. Where FDA recognises a consensus standard, you can declare conformity to it.

Decide whether to ask FDA first

A Pre-Submission fits when the predicate, the testing approach or a difference in technology is uncertain. Send specific questions. The answers are not binding but they are the best signal of what the reviewer will expect.

Prepare in eSTAR and keep the file consistent

The labelling, the comparison table, the test reports and the 510(k) summary must say the same thing. Connect your risk file to the tests. See the ISO 14971 guide.

Testing

Testing Commonly Needed for Hardware Devices

If your devicePlan forWhere to look
Touches the patientBiocompatibility evaluation based on contact type and durationFDA’s guidance on the use of ISO 10993-1
Is supplied sterileSterilisation validation, residuals where relevant, packaging and shelf-life evidenceFDA guidance on sterility information in premarket submissions, and the standards it recognises
Is reusableValidated cleaning, disinfection or sterilisation instructions and the data behind themFDA guidance on reprocessing medical devices
Is electricalElectrical safety and electromagnetic compatibility testingIEC 60601-1 and its collateral standards, as FDA recognises them
Has performance claimsBench testing that matches the claims and the predicate’s performanceThe special controls or guidance for your product code
Is used by lay users or in high-risk tasksHuman factors and usability evidence tied to the use-related risksFDA human factors guidance and IEC 62366-1

A starting list, not a test plan. The classification regulation, product code guidance and the predicate’s summary decide what your device needs. Confirm with FDA guidance current at the time you plan.

Common Mistakes to Avoid

Picking the predicate last. The predicate decides the testing and the comparison. Choosing it after the design is frozen can leave you with no workable predicate.

Letting the indications drift. A new clinical claim or user group can remove your predicate. Settle the wording before you plan the tests.

Testing the wrong configuration. Test reports must cover the final design, materials and packaging. A late change can invalidate the evidence.

Skipping the labelling review. The labelling must support the intended use and match the predicate comparison. Inconsistent claims draw additional information requests.

Under-answering additional information requests. A partial reply can lead to another hold. Answer every point and attach the evidence.

Forgetting what comes after. Clearance is followed by establishment registration, device listing, UDI, complaint handling and device reporting under the QMSR. See the QMSR guide and the UDI guide.

Set Expectations

What a 510(k) Clearance Does Not Do

  • Approve your device — FDA clears a 510(k) on substantial equivalence. Approval refers to a PMA. Do not describe a cleared device as FDA approved.
  • Cover later changes — Changes that could significantly affect safety or effectiveness, or that change the intended use, can need a new 510(k).
  • Open other markets — The EU, UK, Canada and others need their own routes. See the FDA vs EU MDR comparison.
  • Come with a promised date — No expert can guarantee a clearance date, because the review depends on the quality of the submission and FDA’s questions.
FAQ

Frequently Asked Questions

When do I need a 510(k)?

When you intend to market a device in the United States that is not exempt from premarket notification and does not need a PMA. You show that it is substantially equivalent to a legally marketed predicate device. Check the classification regulation and product code for your device first, because many low-risk devices are exempt and some novel devices need a De Novo request instead.

Do I need clinical data for a 510(k)?

Often not. Many 510(k)s rely on bench, biocompatibility and other non-clinical testing. Clinical data may be needed where the technological differences or the indications raise new questions that non-clinical data cannot answer. A Pre-Submission is the way to ask FDA before you invest in a study.

How much does a 510(k) cost in FY 2027?

The FDA user fee for a standard 510(k) is $28,653, or $7,163 for a business with approved small business status, for submissions in the fiscal year from 1 October 2026 to 30 September 2027. The annual establishment registration fee is separate and does not get the small business discount. Testing, consulting and your own time cost more than the fee.

How long does a 510(k) take?

FDA’s goal is a decision within 90 FDA days for most submissions, with a substantive interaction by day 60. FDA days exclude the time the submission is on hold while you answer additional information requests, so elapsed time is usually longer. Complete submissions with the right testing move fastest.

Do I need a new 510(k) when I change a cleared device?

Sometimes. FDA expects a new 510(k) when a change could significantly affect safety or effectiveness, or when there is a major change to the intended use. FDA’s guidance on when to submit a 510(k) for a change to an existing device sets out the decision process. Document the assessment even when you conclude no new 510(k) is needed.

Planning a 510(k)?

Share the device, the likely predicate and your stage. Matching US regulatory experts are put forward after team review.

Last updated: